Features & Benefits:
Advanced Detector Design: The number of detectors, angular range, and layout each contributes to overall system performance.
Superior Instrument-to-Instrument Precision: To provide the same experience regardless of manufacture date, operator skill, or geographic location. Achieve unmatched instrument agreement without the hassle of correlation.
Automatic Laser Alignment in Seconds: Always make perfect measurements with computer-controlled laser alignment. The alignment process is finished in only a few seconds with HORIBA’s innovative approach.
Guaranteed Accuracy and Precision: The LA-960V2 is a highly refined particle size analyzer capable of accurately measuring NIST-traceable size standards within 0.6% of the specification. Fully compliant with ISO 13320 recommendations regarding the measurement of materials on the D10, D50, and D90.
State of the Art Sampling Systems.
Technical Specifications:
Measurement range : 10 nm – 5000 μm.
Measurement Time : Typical measurement takes 60 seconds from liquid filling, sampling and measurement to rinsing.
Measurement Method: Circulation measurement or fraction cell measurement (Fraction cell is optional).
Sample Quantity: Approximately 10 mg – 5 g (Depending on the particle size, distribution and density)
Dispersing Volume: Approximately 180 mL for standard pumping system, 5/10/15 mL for FractionCell accessory, Manual filling : 35 mL, Automatic filling : 40 mL for MiniFlow accessory. Approximately 1 L of LiterFlow option.
Available carrier fluid: Aqua* (A type), Organic solvent (S type) (*Ethanol can be used as a dispersing additive)
Communication: USB 2.0.
Power : AC 100-240V 50/60Hz, 300VA.
Dimensions: 705 (W) × 565 (D) × 500 (H) mm.